PCE-123
calibrator
indicates the nominal value for
simulations and measurement of input
electrical signals such as
frequency and temperature |
|
The PCE-123 process
calibrator / nominal value indicator is a device that does need
a mains connector and that is used to simulate regulation
signals or units en the MSR technique with which it can test and
calibrate almost any unit parameter and measuring instrument.
The nominal value indicator provides an electrical output signal
(mA, mV, V and frequency in Hz as well as temperature in °C
or °F). The calibrator other automatic graduated functions, and
selection of output signal jumps by percentage
(0 to 100%) via the
up/down keys and direct display of the output value.
The device is easy to use and is operated via the keypad. THis
device is attractive due to its high level of accuracy and its
universal range of applications which no electrician or MSR
department should go without, as well as within the industrial
sector.
This device we have on offer is a high quality alternative to
those which are normally found on the market. |
|
- 4 to 20mA (1kΩ,
24V power)
- 0 to 100.00 mV / 0 to 1,0000V / 0 to 12,000V
- type-K, J, E, T temperature sensors (°C or °F)
- Frequency range 1 to 62,500Hz
- Accuracy of 0.025%
- Operated via keypad
- Graduated automatic functions
- Selectable between 0 to 20mA/ 0 to 24mA
- Input 0 to 100% (mA, mV, V )
- Alarm beep 'open' (mA) or 'short circuit' (mV, V
View or print the user's
manual
 |
PCE-123 process calibrator
|
|
Areas of application |
|

|
- The first image shown is the PCE-123 calibrating our SLT
acoustic installation. The measurement range is from 30 to 130dB (A)
for an input signal of 4 to 20mA.
The calibrator emits, in this instance, as signal adjusted
to 50% = 12mA (range 4 - 20mA) that corresponds to an
indicator of 50% = 65 (64.9) dB(A) of the unit indicated.
The accuracy of the SLT can be seen. |
|
- The second image shows an example of us testing the display of
our high precision PCE-T395 thermometer. The calibrator is
adjusted to type-k and with a thermal tension /temperature of 200°C.
The PCE-T395 shows the value accurately and with that the
device is ready to be sent to the client.
- If you wish to learn more about our PCE-T395 thermometer with data
logger function, software and data cable, click
here. |
 |
|
|
- In the last example we test the accuracy of a frequency
measurement with our PCE-DM22 multi-meter with integrated data
logger.
The process calibrator is adjusted to an input signal of
20kHz and as it was shown on our multimeter.
- If you do not have a high spec multi-meter like this in your
shop or in your production process, click
here.
The above examples show a
wide range of uses, in profession practice, for the PCE-123
process calibrator.
The PCE-123 calibrator is highly useful in laboratories, in MSR
workshops or to test measurement instruments in research and
development. |
|
Technical specifications |
|
mA DC current
(maximum load at 1kΩ,
powered by 24V) |
|
Range |
4 to 20mA / 0 to 20mA / 0 to 24mA |
|
Resolution |
1µA |
|
Accuracy |
±0.025% ±3µA |
|
Alert |
beeps when there is a short circuit or
ouput current >1mA . |
|
mV current, V DC
(power 1mA) |
|
Ranges |
0 to 100.00 mV
0 to 10.000 V
0 to 1.0000 V |
|
Resolution |
10 µV
1 mV
1.0000 V |
|
Accuracy |
± 0.05% ±30 µV
± 0.05% ±3 mV
± 0.05% ±300 µV |
|
Alert |
beeps when there is a short circuit or
ouput current >10mV. |
Temperature sensor
type-K, J, E, T
(1 kΩ min.) |
|
Ranges
|
K: -200 to 0°C and 0 to 1370°C
J: -100 to 0°C and 0 to 760°C
E: -10to 0°C and 0 to 700°C
T: -200 to 0°C and T: 0 to 400°C
|
|
Resolution
|
1°C (across entire range)
|
|
Accuracy
|
K: ±1.1°C and ±0.8°C
J: ±0.9°C and ±0.7°C
E: ±0.9°C and ±0.7°C
T: ±1.0°C and ±0.8°C
|
|
Frequency
|
|
Range
|
1 to 125Hz and 126 to 62.5kHz
|
|
Resolution
|
1 to 125Hz / 1Hz 126 to 62.5kHz / 604
readings
|
|
Accuracy
|
±0.04Hz
|
|
Selection
|
604 frequencies
|
|
General characteristics
|
|
Display
|
5 position LCD
|
|
Operating conditions
|
0 to 50°C; <85% r.h. |
|
Storage conditions
|
-20 to 60°C; <85% r.h. |
|
Power
|
1 9V battery or 6 x 1.5V batteries
|
|
Consumption
|
60mA to 180mA (depending on output)
|
|
Battery indicator
|
5.5V for 150mA
|
|
Dimensions
|
88 x 168 x 26mm
|
|
Weight
|
330g
|
|
Contents
PCE-123 process calibrator, carrying case,
type-k adaptor, 6 batteries, test cables with alligator clips and
user's manual |
|
Optional accessories
- ISO calibration certificate for companies that wish to
incorporate the device into their set of internal quality
control tools or for annual recalibrations.
The ISO certification contains a laboratory calibration
with a certificate containing all the measurement data. |
|
 |
 |
|
Here you will find an overview of all the measuring instruments available at PCE Group. |
If you have any doubt, call our
technicians on:
Germany:
+49 (0)29 03 976 99-52
Italy:
+39 0583 975 114
Spain:
+34 967
543 548
|
|
This
page in German
in Italian
in Spanish
in Croatian
in French
 |