PCE-IT100
insulation meter
for insulation resistance up to 4000 MΩ / CAT
III 1000 V / easy to use/
for simple or continuous measurements / max. and min. functions.
|
|
The PCE-IT100 insulation meter detects auxiliary
voltages of 125 V, 250
V, 500 and 1000 V up to a maximum of 4000 MΩ. The PCE-IT100 insulation
meter also offers the ability to measure alternate and continuous
currents up to a maximum 750V / 1000V and resistance up to 400 Ω. It includes test current and a
warning whistle. The PCE-IT100 insulation meter has Data-Hold function to
freeze the current value on the display as well as to hold max. and min.
values while voltage is measured in ohms. This
link shows an overview of all the insulation meters we have available to suit your
needs.
|
|
- It has a large, backlight
display.
- Comes ready to use in its own carrying
case.
- Individual or prolonged measurements can
be
made by pressing the "LOCK" key.
- Auto shut of to protect battery
power.
- Over-range and battery level
indication.
- Optional ISO calibration certificate.
- Security: EN 61010-1, CAT III 1000
V
View /
print user’s manual of the
PCE-IT100 insulation meter.
|

|
|
Technical
specifications
|
|
Measurement
ranges: Voltage AC
Voltage DC
Resistance
Insulation resistance
|
750 V
1000 V
40 Ω
/ 400 Ω
4 MΩ
/ 40 MΩ
/ 400 MΩ
/ 4000 MΩ
|
|
Resolution
Voltage AC
Voltage DC
Resistance
Insulation
resistance
|
1 V
1 V
0,01 Ω
/ 0,1 Ω
0,001 MΩ
/ 0,01 MΩ
/ 0,1 MΩ
/ 1 MΩ
|
|
Accuracy
Voltage AC
Voltage DC
Resistance
Insulation resistance
125 V
250 V
500 V
1000 V
|
±1,2
% ±10 digits
±0,8 % ± 3 digits
±1,2 % ± 3 digits
±(2
% + 10 digits) / ±(2 % + 10 digits) /
±(4 % + 5 digits) / ±(5 % + 5 digits)
±(2
% + 10 digits) / ±(2 % + 10 digits) /
±(3 % + 5 digits) / ±(4 % + 5 digits)
±(2
% + 10 digits) / ±(2 % + 10 digits) /
±(4 % + 5 digits) / ±(4 % + 5 digits)
±(3
% + 10 digits) / ±(2 % + 10 digits) /
±(2 % + 5 digits) / ±(4 % + 5 digits)
|
|
Continuity
test
|
whistle: ≤35 Ω,
test current ≤200mA
|
|
Power
|
6 x 1,5 V AA batteries
|
|
Operating
conditions
|
0 ...
40 ºC
/ <80 % H.r. (working)
-10 ... 60 °ºC / <70 % H.r. (storing)
|
|
Dimensions
|
200 mm
x
92 mm
x
50 mm
|
|
Weight
|
700 g
(con batería)
|
|
Standards
|
IEC10101,
CAT III 1000 V
|
|
Delivery
contents
1 x PCE-IT100 insulation meter, 1 set of test cables, 1 extractable clamp
with
1 m
cable, 6 x batteries, 1 x hand belt, carrying case and user’s manual.
|
|

Test cables are included
with the
PCE-IT100 insulation meter.
|

The
PCE-IT100 insulation meter is
delivered with a robust carrying case.
|
|
Optional
accessories
- Calibration certificate, meeting with DIN ISO 9000. The certification
includes a laboratory calibration and a test certificate containing all the
measurement values.
- Resistance gauge to check the insulation meter.
|
|
Value limits to consider when measuring
insulation
-
for new installations (dry spaces): 1000 Ω x V of the operating voltage (eg.
230V > 230kΩ)
-
for new installations (humid spaces, open air): 500 Ω x V of
the operating voltage
-
grounded device: 1000 Ω
x V of operating voltage
-
device with insulation protector: 2 MΩ
-
electrical tools: 2 MΩ
-
low voltage: 250 kΩ
-
electrical meters: 5 MΩ
The values of test voltages are usually 100V for installations with
low voltages and telecommunication installations, 500V for installations in
buildings and houses and 1000V for low voltage engines and transformers.
|
 |
 |
|
Here you will find an overview of all the
measuring instruments available at PCE Group. |
|
Contact:
Companies House England
Tursdale Business Park
Co. Durham - DH6 5PG
United Kingdom
Phone. +44 ( 0 ) 191 377 3398
Fax +44 ( 0 ) 191 377 3357 |
|
This
page in German
in Italian
in Spanish
in Croatian
in French
 |